Service
|
Description
|
Full Service Inspection (Full Service Inspection Report) |
- Verification of start up process
- Verification of the image uniformity and artifacts
- Verification of temperature measurement against non-calibrated blackbodies
- Visual inspection of exterior parts
- Verification of LCD and viewfinder
- Verification of buttons and joystick
- Verification of mechanical articulation
- Verification of lens operation
- Verification of manual focus and auto focus
- Verification of shutter and filters
- Verification of laser aligment
- Verification of image fusion aligment
- Verification of visual camera
- Verification of GPS functionality
- Verification of Wi-fi and Bluetooth connection
- Verification of PCB embeded battery
- Verification of cooler cooldown time
- Completion of Full Service Inspections report
|
FLIR Factory Certificate of Calibration |
The Certificate of Calibration includes measurement data, as well as traceability information, certificate unique identifier, and measurement uncertainties. |
Firmware Update |
Update of the following firmware* to the latest version available
- OS image kit
- App kit
- Prod kit
- Config kit
- Prod.spc
- GigE drivers
*Specialty products may contain other files systems that require update |
Full Service Preventive Maintenance |
- Adjustment of mechanical articulations
- Adjustment of auto focus
- Adjustment of laser aligment
- Adjustment of fusion aligment
- Tightening all PCB connections and flexcables
- Cleaning of lenses
- Cleaning of the detector window
- Cleaning of product exterior
- Complete Maintenance and Repair report
|
Camera Performance and Image Optimization |
For each lens these are the items included:
- Testing of image uniformization
- Adjustment of the gain and offset of the whole array of pixels
- Replacement of the signal of pixels that are not performing as expected
- Retesting the image uniformization with non-calibrated blackbodies
|
Measurement Adjustment |
For each lens these are the items included:
- Adjustment of all temperature ranges against calibrated blackbodies
- Testing all temperature ranges at 3 levels (range limits and center) against calibrated blackbodies
|